Calibration; Calibration Option 2, Code Kf1 - HBM T10FS Mounting Instructions

Torque flange
Hide thumbs Also See for T10FS:
Table of Contents

Advertisement

40
6

Calibration

T10FS torque flanges deliver an electrical calibration signal that can be
switched at the amplifier end for measurement chains with HBM components.
The measurement flange generates a calibration signal of about 50 % of the
nominal (rated) torque. The precise value is specified on the identification
plate. Adjust the amplifier output signal to the calibration signal supplied by the
connected torque flange to adapt the amplifier to the measurement flange. To
obtain stable conditions, the calibration signal should only be activated once
the transducer has been warming up for 15 minutes.
NOTE
The measurement flange should not be under load when the calibration
signal is being measured, since the calibration signal is mixed additively.
CAUTION
To maintain measurement accuracy, the calibration signal should be
connected for no more than 5 minutes. A similar period is then needed as
a cooling phase before triggering the calibration signal again.

6.1 Calibration Option 2, code KF1

Increasing the excitation voltage from 54 V
triggers the calibration signal.
HBM
to 80 V
(pins 6 and 7, plug 1),
PP
PP
T10FS
A0785 14.0 en