Setting Up The Zero Point - HBM T10FS Mounting Instructions

Torque flange
Hide thumbs Also See for T10FS:
Table of Contents

Advertisement

44
Fig. 7.3: Switch for changing the frequency output voltage

7.3 Setting up the zero point

In the case of the torque flange with the voltage output option (SU2), you can
access two potentiometers by removing the stator cover. You can use the zero
point potentiometer to correct certain zero point deviations. The balancing
range is a minimum of "400 mV at nominal (rated) gain. The end point
potentiometer is used for compensation at the factory and is capped with
varnish so that it cannot be turned unintentionally.
CAUTION
Turning the end point potentiometer changes the factory calibration of
the voltage output.
Fig. 7.4: Setting up the voltage output zero point
HBM
5 V
12 V
symmetrical
asymmetrical
Pos. 2
S1
Pos. 1
S2
End point
Zero point
A0785 14.0 en
T10FS