4.4 Mounting the rotor

Additional installation notes for the speed measuring system can be found in
Section 4.7, Page 23.
NOTE
In general, the rotor identification plate is no longer visible after
mounting. This is why we include with the rotor additional stickers with
the important ratings, which you can attach to the stator or any other
relevant test bench components. You can then refer to them whenever
there is anything you wish to know, such as the calibration signal. To
explicitly assign the data, the identification number and the measuring
range are engraved on the rotor where they can be seen from outside
(see Fig. 4.3).
1. Prior to installation, clean the measurement flange's and counter flanges'
plane surfaces. For safe torque transfer, the surfaces must be clean and
free from grease. Use a piece of cloth or paper soaked in solvent. When
cleaning, make sure that you do not damage the transmitter coils.
Fig. 4.3: Screwed joint of the rotor
HBM
Rotor
Identification number/measuring range
6x or 8x (Z) DIN EN ISO 4762 hexagon socket screw (10.9/12.9)
6x or 8x fastening screw (10.9/12.9)
T10FS
A0785 14.0 en