Option 2, Code Kf1; Adaptation To The Cable Length - HBM T10FS Mounting Instructions

Torque flange
Hide thumbs Also See for T10FS:
Table of Contents

Advertisement

34

5.3 Option 2, code KF1

The stator housing has a 7 pin (Binder 723) male device connector, to which
you link the connection cable for voltage supply and torque signal.
Connector
Binder 723
6
1
5
7
2
3
4
Top view

5.3.1 Adaptation to the cable length

The transmission method between the rotor and the stator determines the
function of the torque flange, which is dependent on:
• the installation situation (for example, covering, area free of metal parts)
• the length of the cable
• the tolerances of the excitation voltage supply
To allow for adaptation to the various conditions, there are three switches in the
stator housing, which can be accessed by removing the stator cover (see
Fig. 5.1).
HBM
Pin assignment
Binder
pin
1
Zero operating voltage
2
No function
3
Pre supply voltage (+15 V)
4
Torque measurement signal (12 V
5...15 kHz)
5
No function
6
Rotor excitation voltage (54 V/80 V
approx.15 kHz)
7
Rotor excitation voltage (0 V)
Shielding connected to enclosure
ground
T10FS
Wire
MS3106
color
connect
or
pin
wh
A
bk
B
bu
C
;
rd
D
PP
;
gn
F
PP
gy
G
A0785 14.0 en