Conditions On Site; Mounting Position - HBM T10FS Mounting Instructions

Torque flange
Hide thumbs Also See for T10FS:
Table of Contents

Advertisement

14

4.1 Conditions on site

T10FS torque flanges are protected to IP54 according to EN 60529. They
must be protected against coarse dirt particles, dust, oil, solvents and humidity.
During operation, the prevailing safety regulations for the security of personnel
must be observed (see "Safety instructions").
There is wide ranging compensation for the effects of temperature on the
output and zero signals of the T10FS torque flange (see specifications on page
64). This compensation is carried out at static temperatures. This guarantees
that the circumstances can be reproduced and the properties of the transducer
can be reconstructed at any time.
If there are no static temperature ratios, for example, because of the
temperature differences between the measuring body and the flange, the
values given in the specifications can be exceeded. Then for accurate
measurements, static temperature ratios must be obtained by cooling or
heating depending on the application. As an alternative, check thermal
decoupling, by means of heat radiating elements such as multiple disc
couplings.

4.2 Mounting position

The measurement flange can be mounted in any position. With clockwise
torque, the output frequency is 10 kHz to 15 kHz. With HBM amplifiers or with
the "voltage output" option, a positive output signal (0 V to +10 V) is present.
In the case of the speed measuring system, an arrow is attached to the head of
the sensor to clearly define the direction of rotation. If the measurement flange
moves in the direction of the arrow, the connected HBM measuring amplifiers
give off a positive output signal (0 V...+10 V).
HBM
T10FS
A0785 14.0 en