4.4 Mounting the rotor

Additional installation notes for the speed measuring system can be found in
Chapter 4.7, Page 24.
NOTE
Usually the rotor identification plate is no longer visible after
installation. This is why we include with the rotor additional stickers
with the important ratings, which you can attach to the stator or any
other relevant test‐bench components. You can then refer to them
whenever there is anything you wish to know, such as the calibration
signal. To explicitly assign the data, the identification number and the
measuring range are engraved on the rotor where they can be seen from
outside (see Fig. 4.3).
1. Prior to installation, clean the plane surfaces of the measurement flanges
and counter flanges. For safe torque transfer, the surfaces must be clean
and free from grease. Use a piece of cloth or paper soaked in solvent.
When cleaning, make sure that you do not damage the transmitter coils.
Fig. 4.3: Screwed rotor joint
HBM
Rotor
Identification number/measuring range
6x or 8x (Z) DIN EN ISO 4762 hexagon socket screw (10.9/12.9)
6x or 8x fastening screw (10.9/12.9)