Make sure that the correct rack is used for the carrier.
If a carrier is damaged, replace it immediately.
Carrier ID
Each carrier ID must be unique.
WARNING
Wrong identification of a carrier
If carriers are not placed correctly on the worktable and if there are unfavorable circum-
stances (barcode labels not within specified limit, distance incorrectly placed carrier to the
barcode reader still enables reading), the barcode reader might read the wrong carrier.
•
When loading carriers, always slide them all the way to the stop at the lock pin.
When carriers need to be removed, always remove them completely from the workta-
•
ble.
8.3.4.2
Racks and Containers
WARNING
If a rack is damaged, replace it immediately.
Make sure that the correct barcode is used for the rack.
ATTENTION
Microplates
Microplates must be positioned correctly on the carrier, seating well in their holder. Make
sure that the microplate does not rest on the holder rim in a slanting position.
ATTENTION
Racks for Disposable Tips
Before positioning the new racks in the DiTi Carrier on the worktable, carefully check the
DiTis for transport or storage damage (refer to the instructions enclosed in the outer pack-
age):
The DiTis must not be damaged
•
The DiTis must not be bent
•
Ensure that the DiTis to be loaded onto the carriers correspond to the ones indicated in
the application software (size, with filter/without filter).
ATTENTION
Crash or erroneous pipetting results when the wrong tips are loaded on the worktable.
•
If tips are longer than expected:
PyroTec™ PRO Hardw are User's Manual, V1.2
Crash of the tips with the labware.
Return to TOC
72
RT-MN024-1.2 02/21