9.4 Precision and Function Tests
9.4.1 Liquid Handling Performance Verification Testing
QC Kit
Lonza recommends the use of the QC kit for liquid handling performance verification test-
ing at least once per year during Preventative Maintenance by a Lonza approved Field Ap-
plication Specialist; a higher frequency may apply, based on the Quality & Regulatory re-
quirements of the laboratory using the PyroTec™ PRO platform.
The QC Kit is based on Artel's patented Ratiometric™ Photometric technology, and is being
made available for Lonza's customers.
Customer benefits include:
•
Traceability of test results to the international standards.
•
Robustness of the method in a daily lab environment
Ease-of-use
•
Alternative
Alternative methods are:
Methods
Gravimetric test; contact Scientific Support for more information on this method.
Software Used
The "Setup & Service" and application software offer a variety of tests to check individual
for Various
functions of the modules.
Tests
Refer to the "Instrument Software Manual"
Refer to the "PyroTec™ PRO Software User Guide"
PyroTec™ PRO Hardw are User's Manual, V1.2
A
Arm guide
B
Arm guide roller
Arm guide and roller
C
Arm rail
D
Arm
Return to TOC
98
RT-MN024-1.2 02/21
Need help?
Do you have a question about the PyroTec PRO and is the answer not in the manual?