At installation or later movement of the instrument, ensure that it is always possible to un-
plug the mains cable at the instrument.
Uninterrupti-
For an optimal operation of instruments and smooth running of the relevant application,
ble Power
Lonza recommends connecting an online UPS, so that the power supply runs via the UPS
Supply (UPS)
with a filter effect.
UPS recommendations:
UPS Type
UPS Power Output
Recommended minimum UPS
size
Note: A switched UPS type that switches over to batteries only after network breakdown is
not recommended.
For further assistance, contact your site manager or your nearest Lonza representative.
A switched UPS type that switches over to batteries only after network breakdown is not
recommended.
Status Lamp
The status lamp is located above the diluters in the center of the instrument front surface.
It displays the instrument operational states with red and green light that is either contin-
uous or flashing. When the status lamp light is red, an acoustic alarm sounds.
The illuminated area is 540 x 18 mm (21.26 x 0.7 in.).
Environmental
All instruments are intended for indoor operation and storage only. The tables below give
Conditions
an overview.
ATTENTION
Barcodes cannot be read due to the influence of sunlight or other light sources on the bar-
code scanner.
•
Do not expose the instrument to direct sunlight.
Do not install strong light sources that may impair the function of the barcode scanner
•
near the instrument.
Operating
Operating temperature
Conditions
Operating humidity
Operating altitude
PyroTec™ PRO Hardw are User's Manual, V1.2
Short-circuit-proof
1.5 times the average consumption of the equipment.
2500 WATTS
15°C to 32°C (59°F to 90°F)
30% to 80% relative (non-condensing) at 30°C (86°F) or be-
low
Max. 2000 m above sea level
Return to TOC
27
RT-MN024-1.2 02/21