Compact Gas Preparation Unit For So Measurement With Hose (Optional); Diagram Of The Compact Gas Preparation Unit - TESTO 300 M-I Instruction Manual

Table of Contents

Advertisement

for SO
measurement with hose (optional)
2
The compact gas preparation unit and the SO
the testo 300 M-I and testo 300 XL-I flue gas analysers ensuring
efficient and correct measurement of the SO
The system consists of an adapted unit with gas hose, compact gas
preparation unit and non-heated standard flue gas probe or the high-
standard industrial flue gas probe.
Use filter pipes with a coarse filter for dirt ingressed flue gases. Dirt
ingress in the gas paths is thus prevented. The hose is a specially
constructed path in which SO
The miniaturised gas preparation unit is integrated into the end of the
hose. It cools the sample gas using the Peltier cooler very quickly thus
dehumidifying the sample gas. This means that the SO
absorbed by the water (reduction of absorption to a minimum). The
condensate builds up in the housing and is emptied by removing one of
the end pieces. The dry sample gas is filtered through a fine filter which
also serves as a water barrier (automatic emptying).
Peltier
element
moist
and
warm
gas
Condensate
Diagram
of the compact gas preparation unit
The use of a compact gas preparation unit with automatic condensate
emptying (peristaltic hose pump) is required for long-term
measurements lasting several hours if the measurement cannot be
interrupted for the purpose of emptying the condensate.
13
Initial operation
Compact gas preparation
hose ideally complement
2
level in flue gases.
2
absorptions do not occur on account of
2
the high gas velocities.
compact gas preparation unit
Ventilator
Heat sink
Condensate
collection container
cannot be
2
Diagram of the
Dry
and
cold
gas
Fine particle
filter