Manual Emptying Of Condensate - TESTO 300 M-I Instruction Manual

Table of Contents

Advertisement

From the flue gas
probe
Input filter
Filter
tube
Filter fleece
(to protect
pump from
residue)
To analyser
Manual emptying of the condensate
Pull off one of the end pieces to let out the condensate.
Note on changing filter
If dirt can be seen in the filter or pump power has audibly slowed down,
the filter should be changed.
Pull off the filter tube in order to change the filter. Screw the drawing key
into the filter and remove from the filter tube. Press the new filter into
the filter tube firmly. You can order spare material for the filter with the
order no. 0554.3370. Please only use this filter.
The condensate can only be emptied if the pump is switched off.
The compact gas preparation is designed in such a way that there
is a certain flow direction. This is marked with arrows on the
housing section. If gas flows in the opposite direction the
Drawing
When assembling the compact gas preparation, ensure that the
screw
seal rings are not damaged.
Output filter
compact gas preparation unit cannot function properly and
this could lead to the instrument breaking down.
53
Maintenance