Differential Temperature Measurement (Testo 300 Xl-I Only); Ambient Co Measurement (Testo 300 Xl-I Only) - TESTO 300 M-I Instruction Manual

Table of Contents

Advertisement

Differential temperature measurement (testo 300 XL-I only)

Only for measuring temperature e.g. flow/return temperature
Temperature 2
Temperature 1
For parallel measurement of the CO level in ambient air during a flue
gas measurement.
Note
Parallel or subsequent measurement of AT is not possible.
The probe should be connected before testo 300 XL is switched on.
The protective cap should be closed during the initialisation phase
A smoker´s breath influences the measurement by approx. 5 ppm.
The CO probe is initialised together with the gas measuring cells
in the testo 300 XL-I analyser.
a. Open protective cap.
b. Hang probe, for example, in your shirt pocket.
The direction in which the gases flow onto the probe influences the
accuracy of the measurement. Optimum measured results are attained
by gently moving the probe back and forth. If the gases flow head-on
onto the sensor the values measured will be too high.
Measuring flue gas (START). The ambient CO value (in ppm only) is
measured parallel to the flue gas values. The ambient CO value is also
saved when the measured values are saved.
Close protective cap.
Measurement menu
N O N A M E
A T !
F F T T
S S T T A A R R T T
M E M O R Y
Differential temperature is taken
over in the flue gas values.
(otherwise the measurements may be incorrect).
The protection cap should only be opened for the duration of the
measurement, close cap straight after measurement (mechanical
protection of sensor and adherence to accuracies).
Cigarette smoke influences the measurement (min. 50 ppm).
Initialising phase (60 s)
Warning if ambient CO value > 50 ppm CO
(limit can be changed in the menu, "Service
25
Example of measurement
A T !
DP
ZOOM
Ambient CO measurement
(testo 300 XL-I only)
Cutoff threshold").
N O N A M E
F F T T
DT
D D i i f f f f . . t t e e m m p p . .

Advertisement

Table of Contents
loading

This manual is also suitable for:

300 xl-i