Flue Gas Measurement (°C, O , Co ) And Co/No/So Measurement - TESTO 300 M-I Instruction Manual

Table of Contents

Advertisement

Measurement menu
AT!
F F T T
0 0
S S T T A A R R T T
MEMORY
N O N A M E
AT!
F F T T
MEMORY
N O N A M E
AT!
MEMORY
Flue gas measurement (°C, O
1. Start measurement by pressing START.
2. Place flue gas probe in the centre of flow:
Search for the max. flue gas temperature FT and then fix the probe
with the cone in this position.
If the flue gas probe is in a vertical position or at an angle of 45° the
condensate drops which run down could result in a sudden drop in
the temperature.
Remedy:
Secure the flue gas probe as shown opposite.
3. You can call up the various measured values via the arrow buttons
and
4. If the measured values do not change, stop the pump.
Place flue gas probe in fresh air.
The measured values remain until you start the pump again by
pressing
S S T T A A R R T T
.
After every measurement rinse the measuring cells with fresh
air (Pump Start/Stop) until the O 2 level is over 20.0 %
and CO and NO levels are below 50 ppm.
Note on SO2 measurement
- Use probe with special hose for SO
measurement)
- Use gas drier for long-term measurement.
- Special filter pipe for SO
Ordering data on page 68 "Upgrade for flue gas probes to SO
18
Measurement example
and CO/NO/SO
measurement (short-term
2
measurement.
2
, CO
)
2
2
measurement
2
."
2