7.7 Replace fuses
Materials and tools required
•
ESD workstation
•
Phillips screwdriver, size 2
•
Ring spanner WAF 14 mm
•
Open-ended spanner WAF 15 mm
•
Pointed pliers
7.7.1 Replace fuse F1
1. Open the device (see "7.2 Open the device"
on page 22).
2. Remove Capno connection adapter
"7.4 Replace the Capno connecting adapter"
on page 24).
3. Lay the device on its rear.
4. Take defective fuse F1 out of the holder using
pointed pliers.
5. Insert a new fuse F1 in the holder using pointed
pliers.
6. Refit Capno connection adapter
7. Close the device (see "7.3 Close the device"
on page 23).
8. Test the device (see "4. Test the device" on
page 8).
7.7.2 Replace fuse F2
1. Open the device (see "7.2 Open the device"
on page 22).
2. Take defective fuse F2 out of the holder using
pointed pliers.
3. Insert a new fuse F2 in the holder using pointed
pliers.
4. Close the device (see "7.3 Close the device"
on page 23).
5. Test the device (see "4. Test the device" on
page 8).