8.7 Exchanging the capacitor

8.8 Replace the Real Time Clock batteries

34
Danger of death! Danger of electric shock! The condenser
must be completely discharged (see Chapter "8.3",
Page 25).
1. Discharge the capacitor and open the device (see
Chapter "8.3", Page 25).
2. Pull the capacitor cable
circuit board with the aid of high-voltage
27
protected needle nose pliers.
3. Remove the capacitor
Chapter "5.4", Page 20).
4. Place a new capacitor onto the printed circuit
board.
Make sure that insulation mats 28 are lying
between the capacitor and the main circuit board.
5. Plug the capacitor cable
board.
Make sure that the polarity is correct +/- (see the
labelling on the printed circuit board and on the
capacitor)! If the cables are mixed up then the upper and
lower device parts will not fit to each other correctly.
6. Close the device (see Chapter "8.6", Page 28).
Danger of death! Danger of electric shock! The condenser
must be completely discharged (see Chapter "8.3",
Page 25).
1. Discharge the capacitor and open the device (see
Chapter "8.3", Page 25).
2. Push the springs slightly backwards and remove
the battery
34
3. Replace with new (button cell CR2032) battery.
Make sure that the polarity is correct!
4. Close the device (see Chapter "8.6", Page 28).
5. Set the date and time with the help of EasyView
(see chapter "Operation" in the operations
manual for PC software EasyView WM 16880).
Repair information and repair instructions