8.2 Changing fuses

At the rear of the OXYMAT 3 are two fuses of
type T 2.5H / 250V.
1. First switch off the device, then pull out
the mains plug.
2. Squeeze together the tabs on each side
of the fuse holder
out.
3. Change the faulty fuse. You will find a
spare fuse in the fuse holder.
4. Push the holder with the new fuse back
into the device.