Hmi As Diagnostic Indicator - ABB 2600T Series Instruction Manual

266 with foundation fieldbus communication
Table of Contents

Advertisement

38
2 6 0 0 T S E R I E S | P R E S S U R E T R A N S M I T T E R S | O I/ 2 6 6/ F F - E N R E V. E
HMI can be set to work in 4 different modes:
Operating mode
Features
One line
Only one variable with its unit code is displayed
One line and bargraph
One variable with its unit code is displayed and another variable can be selected to be displayed in percentage by the bar-graph
Two lines
Two variables with unit code are displayed together (one for each line)
Two lines and bargraph Two variables with unit code are displayed together (one for each line) and another variable can be selected to be displayed in percentage
by the bargraph
The Mode selection can be done through the remote setting of the HMI_MODE parameter or locally from the HMI menu "Display/
Settings/Mode"

HMI as Diagnostic Indicator

During the normal activity, when the display is working as indicator, it can also display diagnostic info when errors or warning are
detected. In this case a message appears in the low side of the display below the bar-graph. The message is the combination of the
the NAMUR NE107 icon and the string relating the source of the error i.e. the component where the problem has been detected.
NAMUR icons
Description
Error / Failure
Functional check (e.g. during
simulation)
Out of Spec (e.g. Sensor temperature
outside the specs limits)
Maintenance required
Example of "Maintenance / Sensor" and "Failure / Electronics" diagnostic messages
Figure 48: Diagnostic messages
When such messages are displayed, it is also possible to get their detail directly from the display itself by pressing the Button (1).
Source of error
ELECTRONICS
SENSOR
PROCESS
CONFIGURATION