Diagnostics; Diagnostic Error Messages; Flow Measurement; Totalizers - ABB AquaMaster4 Operating Instruction

Electromagnetic flowmeter transmitter
Hide thumbs Also See for AquaMaster4:
Table of Contents

Advertisement

A Q U A M A S T E R 4 | E L E C T R O M A G N E T I C F LO W M E T E R T R A N S M I T T E R | O I/ F E T4 0 0 - E N R E V. D
56

8 Diagnostics

Diagnostic error messages are displayed by pressing the
Diagnostics tab:
Aqm4
ID:4294967295
Last access: 10/05/2021 16:35:18

Flow Measurement

Flow Sensor Communications
Empty Pipe
High Flow Alarm
Low Flow Alarm

Totalizers

Totalizer Reset
Pulse Output Error

Maintenance Operating Modes

Simulation Mode Activate
Calibration Mode Activate
?
Process Values
Configurtion
Figure 77 Example Diagnostics window
Messages displayed are for the last flowmeter selected from
the Instruments List – see page 48.
The following icons indicate the diagnostic status:
Table 6 Diagnostic icons
Icon
Status
Failure
Check
function
i
Diagnostics
Icon
Status
Out of
specification
Maintenance
required

Diagnostic error messages

Flow Measurement
Table 7 Flow Measurement – diagnostic messages
Icon
Corrective action
Message
Remote sensor is not plugged in, or not
connected. If connected, this error indicates
Flow Sensor
a connection problem, with no sensor data
Communications
connection.
Check the sensor terminal box wiring.
The pipe is not full of water. Check the pipe.
An empty pipe is indicated when either, or both
measuring electrodes have resistances which
exceed the resistance trip level.
Check each electrode SigA & SigB Impedance.
Empty Pipe
If one electrode resistance is far higher than the
other, it could indicate a wiring fault.
For very low conductivity application (< 20 uS/
cm) the electrode trip level may need to be
increased to prevent unwanted empty pipe
alarms.
The flowrate has exceed the upper flowrate alarm
High Flow Alarm
trip level. Ensure this alarm threshold is set
correctly. Disable this alarm if not required.
The flowrate is lower than the low flowrate alarm
Low Flow Alarm
trip level. Ensure this alarm threshold is set
correctly. Disable this alarm if not required.
Totalizers
Table 8 Totalizers – diagnostic messages
Icon
Message
Corrective action
Totalizer Reset A user has reset the totalizers.
The pulse output has saturated at the maximum
Pulse Output
output frequency. An external counter will under
Saturation
read compared to the true volume. Reduce the
Warning
Pulses per unit volume on the Pulse Output
Setting menu to lower the output frequency.
Maintenance Operating Modes
Table 9 Maintenance Operating Modes – diagnostic messages
Icon
Message
Corrective action
The user has put the flowmeter into simulation
Simulation
mode. The displays and outputs are set based on
Mode Active
the entered simulated input value. This error is
cleared by turning off simulation mode.
The user has put the flowmeter into calibration
Calibration
mode. This error is cleared by turning off
Mode Active
calibration mode.

Advertisement

Table of Contents
loading

Table of Contents