. . .Device Diagnostic Mapping
Cat .
Error
Electronic Interface error.
Non-Volatile Sensor memory
burn error
Non-Volatile Electronics
memory burn error
Replace Info
PILD Output
PILD Changed Operating
Conditions
RB_fail / maintenance / offspec / check_
ACTIVE (indexes 47, 48, 49, 50)
Exchange of non-critical data between sensor and
electronics is precluded due to problem in the
transmitter circuit of the electronics or in receiver
Bit 14
circuit of the sensor
Writings to the Sensor non-Volatile Memory were
not successful. The device works without problems
but any replacement operation is compromised
Bit 15
because the back-up configuration is not updated
Writings to the electronic Non-Volatile Memory
were not successful. The device continue to work
without problems but after the next power cycle
Bit 16
the last configuration will be lost
An element of the transmitter has been changed
Bit 17
(sensor or electronics) and the correct replacement
operation must be executed
The PILD algorithm has detected impulse lines
Bit 18
plugged. The Plugged Line can be one among:
The pressure value currently detected is too
Bit 19
different from what used for the PILD Training
RB_EXTENDED_ACTIVE (index 69)
The Replace operation is required after the
changing of the electronics or of the sensor.
Bit 5
Both the directions are allowed, from
Electronic (CB) to Sensor (FE) or from Sensor
(FE) to Electronic (CB)
The Replace operation is required after the
changing of the electronics or of the sensor
Bit 6
but it is allowed only from Electronic (CB) to
Sensor (FE)
The Replace operation has been attempted
Bit 7
but with wrong direction
PILD algorithm has detected both impulse
Bit 10
lines plugged.
PILD algorithm has detected a plugged
Bit 11
impulse line on the HIGH side.
PILD algorithm has detected a plugged
Bit 12
impulse line on the LOW side.
PILD algorithm has detected one plugged
Bit 13
impulse line.