Micromeritics Gemini VII 2390 Confirm Installation Instructions And Checklist page 57

Hide thumbs Also See for Gemini VII 2390 Confirm:
Table of Contents

Advertisement

Gemini VII 2390 Confirm
f. Click
g. When pressure stabilizes, click
or type in the pressure reading from the pressure calibration standard.
h. Press
i. Click
j. Enable manual control and open valves 5-b and 5-s until the sample pressure is
between 0 and 300 mmHg. Close all valves.
k. Repeat step 2, but change the servo setting in step 2e to 380 mmHg.
l. Proceed if the values for the transducers and pressure gauge are within 2 mmHg of 380
mmHg. If the values are not within this range, repeat the calibration.
Feb 09
. The following dialog is displayed.
Next
.
Enter
when the calibration terminates.
Close
Gemini VII 2390 Confirm Installation Instructions and Checklist
Calibrating Transducers and Servo Valves
to enter the reading from the gauge,
From Reference
51

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Gemini VII 2390 Confirm and is the answer not in the manual?

Questions and answers

Table of Contents