Read this section of the manual before beginning installation of the Model 5000 ana-
lyzer system. Failure to do so may impair the protection against fire, electrical shock,
CAUTION
and personal injury originally provided for by this equipment. In addition, failure to
follow the installation and start-up instructions may void the instrument's warranty.

Prior to Installation

The 5000 Field Unit is approved for the following hazardous area classifications:
Class I; Group B, C, D; Division 1 as defined by the U.S. National Elec-
trical Code (NEC) and
Zone 1.
The Model 5000 analyzer system is typically installed with the field unit near the sample
tap (normally in a hazardous area) and the 5000 controller located in a control room.
The controller is suitable for installation in NEC Class 1 ; Groups A, B, C & D; Division
2 areas. The field unit and controller may be separated by up to 600 meters (2000 feet).
Prior to installation, a site plan should be prepared defining the mounting locations of
the field unit, controller, and sample system components. The plan should also detail the
AC input(s), interconnect wiring, and sample line routing.

Unpacking and Inspection

Remove components from the packing case(s) carefully and check contents against pack-
ing list. Inspect all components for obvious damage and broken/ loose parts or fittings.
Notify the carrier and AMETEK Service (1-800-222-6789) immediately if parts are miss-
ing or damage is found.