Gas Connections to Analyzer

Proper sample line purge must be performed before connecting the sample line to
!
analyzer. Condensate in the sample line can migrate to the cell which may result in
irreparable damage to the crystal oscillator.
CAUTION
Consult plant safety personnel for appropriate venting guidelines.
NOTE
32 | 5000 Single-Point Moisture Analyzer
1. Open the main process sample line valve and pressure reducer care-
fully. Vent the line to an appropriate area and purge it with clean, dry
instrument air or nitrogen.
2. Close the process sample line valve and pressure reducer and connect
the sample line to the
3. Connect the conditioned reference gas line (from either the sample or
a separate source) to the
4. Connect vent line to analyzer
5. Vent the line to appropriate area.
/
inch sample inlet fitting.
1
8
/
inch reference inlet fitting.
1
8
/
inch exhaust port fitting.
1
4