Field Unit Electronics

Power Distribution

The AC supply is applied through circuit breaker CB1 and thermal
cutout S2 to the primary of transformer T1.
The T1 secondary (35 VAC) is connected directly to rectifier and regu-
lator circuits on the electronic control board.
Diodes CR4 and CR5 supply unregulated +24 volts and feed the
series-connected voltage regulators.
VR1 supplies +12 volts and VR2 supplies +5 volts.
Diodes CR2 and CR3 supply unregulated +24 volts for the cell oven.
CR15 and CR16 supply unregulated +24 volts for the moisture gen-
erator oven.

Power distribution failure symptoms

• Valves do not switch, but output reading could be on display.
• One or more LEDs do not light, depending on faulty circuit.
• No moisture generator heat.
• No oven heat; CR1 and CR14 on steady.

Electronics control board voltage checks

• +5 volts — TP5 J7
• +12 volts — TP4 J6
• +24 volts — TP2 J4 (0 volt — TP1 J3)

Heat Control

There are two, identical, +24-volt heat control circuits: one for the cell oven and one for
the moisture generator. Each has a heater, an inverting amplifier (U1 or U2), a thermistor
(RT1 or RT2), a triac (Q1 or Q2), an LED (CR1 or CR14), a temperature adjustment pot
(R17 or R26), and miscellaneous resistors and capacitors.