System Tests Menu

Numerous self-test routines can be run periodically or used to assist in troubleshooting
the analyzer. Press the Test/Config key and enter the Level 1 or higher password. Scroll
down to System Tests Menu and press Enter. Select a test from the following list and
then follow directions displayed on screen.
Normal serial communications are disabled while system tests are running.
NOTE
% of
Full Scale
-15
-10
-5
0
5
10
15
20
Figure 11.
4 to 20 mA signal for
percentage output.
Test Display
Verifies that each segment of the display is functioning properly by
displaying a checkerboard pattern for several seconds and then re-
versing it. Press Enter to begin test or Cancel to quit.
Test Analog Out
Sets the signal on the analog output to an mA value corresponding to
any percentage of full scale for those outputs. Default percentage is
0% which sets the outputs to 4 mA. A different percentage from –15%
to +100% may be entered at the "New Value?" prompt. Figure 11 lists
the correct value of the 4 to 20 mA signal for various percentages.
Press Enter to begin test or Cancel to quit.
Test Analog Inpts
Verifies that alarm voltage signals from the cell and moisture genera-
tor ovens and process pressure transmitter (if live pressure input is be-
ing used) are present and tracking. Press Enter to begin test or Cancel
to quit. First three lines of the display will appear as follows:
% of
% of
Output (mA)
Full Scale
1.6
25
2.4
30
3.2
35
4.0
40
4.8
45
5.6
50
6.4
55
7.2
60
Output
Full
(mA)
Scale
8.0
65
8.8
70
9.6
75
10.4
80
11.2
85
12.0
90
12.8
95
13.6
100
Essential Health & Safety Requirements
Output
(mA)
14.4
15.2
16.0
16.8
17.6
18.4
19.2
20.0
|
35

Advertisement

Table of Contents
loading

Table of Contents