Warnings

Review and follow these Warnings to avoid personal injury or environmental contamination.
High voltages are present in the analyzer housing. Always disconnect power and/or external
!
power sources to the analyzer before opening any covers or doors on the analyzer to check or
perform maintenance on any components within the enclosures.
WARNING
If it is necessary to open the analyzer's covers or doors while the circuits are live, first monitor
the area for flammable gases (and proceed only when the area is safe).
When the enclosure is open, take appropriate precautions to avoid electrical shock. Hazardous
voltages are present inside.
All electrical connections, adjustments, or servicing of the analyzer should be performed only
by properly trained and qualified personnel.
!
All electrical connections, materials, and methods (plus all safety policies and procedures)
WARNING
must be made in compliance with local wiring regulations and electrical code for the hazardous
area, as specified by the Owner Company, local electrical-inspection authority, and National/
EU regulations.
!
Follow appropriate regulatory and/or company procedures to lock out the analyzer while work-
ing on the analyzer electronics.
WARNING
Before working on the analyzer, test the area for flammable gases. If a flammable gas atmo-
!
sphere is present, do not open any doors without first disconnecting power and/or alternate
power sources to the analyzer.
WARNING
The field unit may contain flammable or toxic process gas. Remove the field unit dome lid in a
!
well-ventilated area only after electrical power is removed and the surrounding are is known to
be nonhazardous..
WARNING
Before working on the sample system, confirm that the system has been purged with Zero fluid
and is isolated (blocked in) from the process.
!
If the
Sample System Enclosure
tions to avoid burning yourself.
WARNING
!
Verify ground continuity of all equipment before applying power.
WARNING
6 | 5000 Moisture Analyzer