Separate Gas Source Used As Reference

Configure sample system as shown in Figure 3-7b. Standard and optional conditioning
components (dryers, contaminant traps) should be strapped in the upright position to
support beams.
Dryer(s) must be vertical to avoid channeling (voids in the sieve).
NOTE
CONTAMINANT
TRAP
(OPTIONAL)
PRESSURE
REDUCER
SHUTOFF
VALVE
REFERENCE
GAS SOURCE
Figure 9.
Sample system using separate gas
source as reference.
A separate source (if present) should also be run through an oil/air separator to re-
move trace quantities of oil from the gas. Oil in the reference stream will cause rapid
contamination of the analyzer cell.
NOTE
CONTAMINANT
TRAP
(OPTIONAL)
MOLECULAR
SUPER-ACTIVATED
SIEVE DRYER
MOLECULAR SIEVE DRYER
(<1 ppm)
(OPTIONAL)
Essential Health & Safety Requirements
SAMPLE
SHUTOFF
VALVE
PRESSURE
SHUTOFF
REDUCER
BYPASS
FLOWMETER
(OPTIONAL)
VENT TO
APPROPRIATE
AREA
REFERENCE
FIELD
SAMPLE
UNIT
PRESSURE
REDUCER
BYPASS
FLOWMETER
(OPTIONAL)
PROBE
SAMPLE
VALVE
SOURCE
TO VENT
PROBE
SHUTOFF
SAMPLE
VALVE
SOURCE
TO VENT
|
31