Conditioned Sample Gas Used As Reference - Ametek 5000 Manual

Moisture analyzer
Table of Contents

Advertisement

For some special applications, we recommend that the sample should also be run
through a contaminant trap and a super-activated dryer. Refer to the supplemental in-
formation supplied with the system to determine if these optional components are to be
used.
Moisture contamination of the dryer(s) can cause incorrect sample moisture data.
To prevent this contamination, never leave either end of a dryer open and exposed to
NOTE
atmospheric moisture.

Conditioned sample gas used as reference

Configure sample system as shown in Figure 3-7a. Standard and optional conditioning
components (dryers, contaminant traps) should be strapped in the upright position to
support beams.
Dryer(s) must be vertical to avoid channeling (voids in the sieve).
NOTE
MOLECULAR
SIEVE DRYER
(<1 ppm)
Figure 8.
Sample system using conditioned
sample gas as reference.
30 | 5000 Single-Point Moisture Analyzer
CONTAMINANT
TRAP
(OPTIONAL)
SUPER-ACTIVATED
MOLECULAR SIEVE DRYER
(OPTIONAL)
CONTAMINANT
TRAP
(OPTIONAL)
MOLECULAR
SUPER-ACTIVATED
SIEVE DRYER
MOLECULAR SIEVE DRYER
(<1 ppm)
(OPTIONAL)
VENT TO
APPROPRIATE
AREA
REFERENCE
FIELD
SAMPLE
UNIT
SAMPLE
SHUTOFF
VALVE
PRESSURE
SHUTOFF
REDUCER
VALVE
BYPASS
FLOWMETER
(OPTIONAL)
TO VENT
VENT TO
APPROPRIATE
AREA
REFERENCE
FIELD
SAMPLE
UNIT
PROBE
SAMPLE
SOURCE

Advertisement

Table of Contents
loading

Table of Contents