multiFLEX
M10 and M5
User Manual
1
1

Advertisement

Table of Contents
loading

Summary of Contents for ProMinent multiFLEX M10

  • Page 1 multiFLEX M10 and M5 User Manual...
  • Page 2 M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 3: Table Of Contents

    CONTENTS Safety 1. STANDARD OPERATIONS What’s Happening Now Checking & Clearing Alarms Checking & Changing Setpoints Calibrating Sensors Changing Biocide Timing Adjusting % Feeds Userid - Passwords 2. ADDITIONAL OPERATIONS Setting Biocide Prebleed-Lockout Adjusting Boiler Blowdown Timing Sensor Diagnostics Control Diagnostics System Diagnostics 3.
  • Page 4 Saftey Electrical Shock Hazard CAUTION: The operator of this instrument is advised that if the equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impaired. Opening the controller enclosure with the controller plugged in, exposes the user to AC line voltages on the lower of the two controller circuit boards.
  • Page 5 W h a t ’ s H a p p e n i n g N o w 1 . 1 System:2003-10-03 Power UP, first display, current date S/N: M0389001 Key ENTER for System menu Alarms: 16:38:11 Key DOWN & enter to clear all alarms D G 2 Sys And to view detail on Sensor ‘D’, Relay ‘2’...
  • Page 6: Checking & Clearing Alarms

    C h e c k i n g & C l e a r i n g A l a r m s 1 . 2 1 o f 3 System:2003-10-03 CHECK ALARMS S/N: M0389001 Power UP, first display Alarms: 16:38:11 Key DOWN to Alarms &...
  • Page 7 C h e c k i n g & C l e a r i n g A l a r m s 1 . 2 2 o f 3 CLEAR ALARMS System:2003-10-03 Power UP, first display S/N: M0389001 Alarms: 16:38:11 Key ENTER to view detail on Sensors ‘C’...
  • Page 8 C h e c k i n g & C l e a r i n g A l a r m s 1 . 2 3 o f 3 RESET RELAY USING CLEAR ALARMS Biocide A ON:owes 16.6 Key UP or DOWN to the output relay Biocide B &...
  • Page 9: Checking & Changing Setpoints

    C h e c k i n g & C h a n g i n g S e t p o i n t s 1 . 3 Tower Conduct'vty 986 uS Key UP or DOWN to the output relay &...
  • Page 10: Calibrating Sensors

    C a l i b r a t i n g S e n s o r s 1 . 4 1 o f 2 SENSORS Tower Conduct'vty 986 uS Key ENTER at selected sensor Bleed Valve OFF: Setpoints Tower Conduct'vty Diagnostics Key DOWN to Calibrate &...
  • Page 11 C a l i b r a t i n g S e n s o r s 1 . 4 2 o f 2 CONTACT HEAD & TURBINE METERS Make-up meter 23400 gal Key ENTER at selected water meter Inhibitor Pump OFF: Setpoints Make-up meter...
  • Page 12: Changing Biocide Timing

    C h a n g i n g B i o c i d e T i m i n g 1 . 5 1 o f 2 Biocide A MODIFY EXISTING EVENT, REVIEW EVENTS 36.5 min Biodispersant Key ENTER at selected chemical pump This one’s powered by Relay No.5 OFF: Setpoints Biodispersant...
  • Page 13 C h a n g i n g B i o c i d e T i m i n g 1 . 5 2 o f 2 Biodispersant ADD AN EVENT OR EVENTS Timed Events..Add an Event Key ENTER at selected chemical pump Events,8 Key DOWN to Timed Events &...
  • Page 14: Adjusting % Feeds

    A d j u s t i n g % F e e d s 1 . 6 1 o f 2 Tower 1 Inhibitor ADJUST BASE FEED 6.2 min Key UP or DOWN to the pump control Tower 2 Inhibitor &...
  • Page 15 A d j u s t i n g % F e e d s 1 . 6 2 o f 2 Tower 1 Inhibitor ADJUST % BLEED FEEDS 6.2 min Tower 2 Inhibitor Key UP or DOWN to the pump control OFF: No Control &...
  • Page 16: 1 . 7 U S E R I D - P A S S W O R D S

    1 . 7 U s e r i d – P a s s w o r d s System: IGNORE IF PASSWORDS NOT ‘ON’ Passwords... IN YOUR CONTROLLER Login Required ‘YES’ appears if passwords are ON’ YES,public Your present USERID also displays. Editing Value: Key Password..
  • Page 17: Setting Biocide Prebleed-Lockout

    S e t t i n g B i o c i d e P r e b l e e d - L o c k o u t 2 . 1 1 o f 2 Biocide A OFF: No Control Key UP or DOWN to the biocide pump control Biocide B &...
  • Page 18 S e t t i n g B i o c i d e P r e b l e e d - L o c k o u t 2 . 1 2 o f 2 Biocide A You can ‘Prebleed’ until this sensor measures Prebleed-Lockout---- the conductivity that you set Prebleed Sensor...
  • Page 19: Adjusting Boiler Blowdown Timing

    A d j u s t i n g B o i l e r B l o w d o w n T i m i n g 2 . 2 1 o f 2 Boiler 1 Cond. Key UP or DOWN to the boiler blowdown control 2546 uS &...
  • Page 20 A d j u s t i n g B o i l e r B l o w d o w n T i m i n g 2 . 2 2 o f 2 If the measured conductivity is below B1 Blowdown the TURN OFF setpoint, Captured Sample-----...
  • Page 21: Sensor Diagnostics

    S e n s o r D i a g n o s t i c s 2 . 3 1 o f 2 Tower Conduct'vty Key UP or DOWN to the desired sensor 986 uS & key ENTER Bleed Valve OFF: Setpoints This example is a cooling tower conductivity sensor Connected to input ‘E’...
  • Page 22 S e n s o r D i a g n o s t i c s 2 . 3 2 o f 2 Tower Conduct'vty Diagnostics..Selecting ‘Reset to Factory’ during sensor calibration Default Gain sets the Gain Multiplier to the Default Gain 5.6000 Tower Conduct'vty Conductivity adjusts Gain Multiplier to calibrate...
  • Page 23: Control Diagnostics

    C o n t r o l D i a g n o s t i c s 2 . 4 1 o f 4 Make-up Meter INHIBITOR FEED EXAMPLE page 1 38400 gal Inhibitor Pump Key UP DOWN to Inhibitor Pump & ENTER. ON: owes The pump is ON and owes 6.1 minutes of ON time Inhibitor Pump...
  • Page 24 C o n t r o l D i a g n o s t i c s 2 . 4 2 o f 4 Inhibitor Pump INHIBITOR FEED EXAMPLE page 2 Diagnostics..Then turn ON for In this example, after each 100 gallons, the Inhibitor pump turns ON for 20 seconds.
  • Page 25 C o n t r o l D i a g n o s t i c s 2 . 4 3 o f 4 Boiler 3 Cond. CAPTURED SAMPLE EXAMPLE page 1 3628 uS B3 Blowdown Key UP DOWN to the blowdown control & ENTER. 1.2 min Blowdown valve is ON and has been ON for 1.2 minutes...
  • Page 26 C o n t r o l D i a g n o s t i c s 2 . 4 4 o f 4 B3 Blowdown CAPTURED SAMPLE EXAMPLE page 2 Diagnostics..Turn OFF setpoint Controller checks Turn OFF Setpoint at the end of every Measure period.
  • Page 27: System Diagnostics

    S y s t e m D i a g n o s t i c s 2 . 5 1 o f 2 System:2003-10-03 Power ON display S/N: M0389001 OR key UP DOWN to System: & ENTER. Alarms: 16:38:11 D G 2 Sys System: Key ENTER at Diagnostics...
  • Page 28 S y s t e m D i a g n o s t i c s 2 . 5 2 o f 2 Current state of alarm contacts System: Diagnostics..The controller may be configured to Alarms OPEN or CLOSE alarm contacts on alarm. CLOSED Displays the current state of the dry contacts at the AL1 &...
  • Page 29: Read This First

    R e a d t h i s f i r s t ! 3 . 1 ONE CONTROLLER – MANY APPLICATIONS Controllers are shipped configured with a wide range of sensors for one or more cooling towers, multiple boilers, hot & chilled closed loops, condensate monitoring, waste water control and monitoring…...
  • Page 30: Connect Sensors

    C o n n e c t S e n s o r s 3 . 2 1 o f 4 M7 Type Controllers Water meters are typically connected to inputs O,P & Q. Flowswitches, fail-to-sample sensors and contact sets connected to R, S & T. M14 Type Controllers Water meters are typically connected to inputs O,P &...
  • Page 31 C o n n e c t S e n s o r s 3 . 2 2 o f 4 Typical Sensor Driver Connections ORP & pH to S+ Black to S- Bare Shield to '-' White to T+ Clear, center wire to '+' Green to T- S+ S- T+ T-...
  • Page 32 C o n n e c t S e n s o r s 3 . 2 3 o f 4 M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 33 Typical 4-20mA Output & Remote Proportional Input Connections Monitor Control Loop Powered 4-20mA Input Pump or Valve +24V IN 1+ 1- 2+ 2- 15 VDC Part: IO Controller DC Supply 15 to 22 VDC Supply Dual 4-20mA Output Loop Powered Dual output driver Sensor shown.
  • Page 34 C o n n e c t S e n s o r s 3 . 2 4 o f 4 M7 Module Sensor Locations M14 Module Sensor Locations M7 Module M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 35: Connect Pumps, Valves, Solenoids

    C o n n e c t P u m p s , V a l v e s , S o l e n o i d s 3 . 3 M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 36 M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 37 V e r i f y C o n t r o l s 3 . 4 1 o f 2 VERIFY SENSORS MATCH CONTROLS Controlling sensors are followed by the Tower Make-up controlled pumps, valve & solenoids. 12650 gal Inhibitor Pump Make-meter connected to input ‘O’.
  • Page 38 V e r i f y C o n t r o l s 3 . 4 2 o f 2 VERIFY – MODIFY SETPOINTS Tower pH 7.62 pH The Tower pH sensor at input ‘C’ Controls the acid pump powered by relay 3 Acid Pump 8.6 min Key ENTER at Acid Pump to verify-modify setpoints...
  • Page 39: Setting Sensor Alarms

    S e t t i n g S e n s o r A l a r m s 3 . 5 Tower pH VERIFY – MODIFY ALARMS 7.62 pH Key UP DOWN to the target sensor Acid Pump and key ENTER 8.6 min Tower pH Diagnostics...
  • Page 40: Setting Output Alarms

    S e t t i n g O u t p u t A l a r m s 3 . 6 VERIFY – MODIFY Tower pH FEED LIMIT ALARMS 7.62 pH Acid Pump Key UP DOWN to the target control and key ENTER 8.6 min Acid Pump...
  • Page 41: Verify Interlocks

    V e r i f y I n t e r l o c k s 3 . 7 Conductivity INTERLOCKS PREVENT CONTROLS 1245 uS FROM TURNING ON Tower Bleed Key UP DOWN to the target control 8.6 min and key ENTER Tower Bleed Diagnostics Configure...
  • Page 42 V e r i f y B l o c k i n g R e l a y s 3 . 8 BLOCKING PREVENTS TWO Inhibitor Pump CONTROLS FROM TURNING ON OFF:Blocked 3 AT THE SAME TIME Oxidizing Biocide This example shows an Inhibitor Pump,1 blocked 14.2 min when the Oxidant Pump,3 is ON...
  • Page 43: Selecting Special Controls

    S e l e c t i n g S p e c i a l C o n t r o l s 3 . 9 Inhibitor Pump This example shows an Inhibitor Pump OFF: No Control with ‘Bleed then Feed’ Special Control Biocide A 8.6 min Key ENTER...
  • Page 44: Modifying Variable Cycles

    M o d i f y i n g V a r i a b l e C y c l e s 3 . 1 0 1 o f 2 Conductivity Variable Cycles may be used where varying 1384 uS make-up conductivity cause water treatment fault.
  • Page 45 3 . 1 0 M o d i f y i n g V a r i a b l e C y c l e s 2 o f 2 Tower Bleed Variable Cycles..Key ENTER to modify High Range High Range 1000 uS When the Make-up Conductivity is...
  • Page 46: Modifying Feed Verification

    M o d i f y i n g F e e d V e r i f i c a t i o n 3 . 1 1 1 o f 2 Tower Make-up Controller Option: Feed Verification calculates 18425 gal Inhibitor ppm and Inhibitor tank level with fail-to-feed alarms.
  • Page 47 3 . 1 1 M o d i f y i n g F e e d V e r i f i c a t i o n 2 o f 2 Select one: Fixed Cycles applies a user set cycles ppm method select...
  • Page 48 T i m e & D a t e 3 . 1 2 Time & Date are battery backed System:2003-10-27 Adjusting Timer & Date may not be S/N: M0389001 required on start-up Alarms: 16:38:11 none Key ENTER at System: System: Enable I/O Configure Key DOWN to Time &...
  • Page 49: Spare Parts

    S p a r e P a r t s 4 . 1 Fusing Protects Rating / Type Manufacturer - Vendor Power Relays Littlelfuse, Type 217, 250VAC Fuse 1-5 6.3 Amps @ 120VAC Digikey Part# F953-ND & 3 Amps @ 250VAC Digikey Part# F950-ND Fuse 6-10 5mm x 20mm,...
  • Page 50: Technical Support

    T e c h n i c a l S u p p o r t 4 . 2 ProMinent Fluid Controls, Inc. (formerly Aquatrac, Inc.) 136 Industry Drive Pittsburgh, PA 15275 412-787-2484 8:00AM - 5:00PM EST Other Keypad Functions Navigation to keypad functions is detailed in Section 4.4, Keypad Navigator...
  • Page 51 S p e c i f i c a t i o n s 4 . 3 1 o f 4 Controllers configured M7-PR5, M7-PR10, M14-PR5 & M14-PR10 Analog – Digital I/O Rating - Detail Notes Analog Inputs M7: 7 Analog Sensors 3 Dual &...
  • Page 52 S p e c i f i c a t i o n s 4 . 3 2 o f 4 Controls Rating - Detail Notes ON/OFF User set deadband and controlling Any relay can be user sensor(s) or contact set. configured for any Control.
  • Page 53 S p e c i f i c a t i o n s 4 . 3 3 o f 4 Controls Rating - Detail Notes Alarms – Feed Limit Timers Minutes / Actuation User defined OFF on Feed Limit Minutes / Day Auto reset on Bleed &...
  • Page 54 S p e c i f i c a t i o n s 4 . 3 4 o f 4 Electrical Rating - Detail Notes AC Input 120 or 240 VAC, 50/60Hz, Switch selectable Fusing M10 Version 5x20mm, 120VAC fusing: 12.6 Amps @ 120VAC Relays 1-5 &...
  • Page 55 Temperature 0C to 40C M714_User_Master.doc Updated Aug. 25, 2015...
  • Page 56: Keypad Navigator

    K e y p a d N a v i g a t o r 1 o f 2 4 . 4 Activity Top Level then then UP DOWN DOWN & Notes & ENTER 4-20mA Output Output Configure ENTER & ENTER to toggle MANUAL-AUTO MANUAL-AUTO C1..C8...
  • Page 57 K e y p a d N a v i g a t o r 2 o f 2 4 . 4 Activity Top Level then then UP DOWN DOWN & Notes & ENTER Enable I/O System: Enable I/O DOWN to inputs or outputs Key ENTER &...
  • Page 58: Revision Log

    R e v i s i o n L o g 1 o f 1 4 . 5 Issued Contents 2004-01-01 Original issued 2004-03-16 Section 1.7. Removes ‘userid’ selection for keypad log-in. Applies only to controllers with Keypad Passwords ON 2004-04-05 Cover page revised, Repagination, Safety grounding information added to page 3.