Connecting the Exhaust
[1]
Connect 7.6 cm (3 in.) ELBOWS and rigid DUCT or flexible EXHAUST HOSE between the processor
EXHAUST PORT and the BUILDING EXHAUST DUCT.
Do not make a solid connection at the BUILDING END. Do not connect the EXHAUST DUCT to
the processor at this time.
[2]
Measure the static pressure in the DUCT using a modified J TUBE (CHECK TUBE 592380) and AIR
METER TL-2431. Make the measurement 30.5 cm (12 in.) from the end of the DUCT to be connected
to the processor.
AIR FLOW
[3]
To obtain the correct static pressure, as shown in Table 1, adjust the clearance at the BUILDING END of
the DUCT as illustrated.
34
NOTE
30.5 cm
(12 in.)
AIR METER
Figure 39 Measuring the Static Pressure
Table 1 Static Pressures
Duct Diameter
7.6 cm (3 in.)
10.2 cm (4 in.)
EXHAUST
HOSE
RUBBER HOSE or
MODIFIED J TUBE
TAPE
CENTER
CONNECTOR
Negative Static Pressure, (Water Head)
MIN
0.76 mm (0.03 in.)
0.25 mm (0.01 in.)
HOSE SUPPORT
H027_0100BCA
H027_0100BA
MAX
1.02 mm (0.04 in.)
0.51 mm (0.02 in.)
636738