Configuring Prism Wash; Wash Cycle; Figure 41 Automatic Prism Wash Cycle - Vaisala PR-23 Series User Manual

Process refractometer
Hide thumbs Also See for PR-23 Series:
Table of Contents

Advertisement

The value of the bias parameter f00 is added to the concentration value:
NEW CONC = OLD CONC + f00.
7.5 Configuring prism wash
In some applications the process flow does not keep the prism clean because of sticky
process medium or low flow velocity. In these applications the prism can be automatically
cleaned by installing a wash system.
The prism wash settings for sensors A and B are independent of each other. The wash
system is active if a relay has been configured to be a wash relay, see
(page
66), and the wash time is not zero. An automatic wash function can be configured so
that both sensors have different parameters.
More information
‣ Prism wash systems (page 39)
7.5.1 Wash cycle
Figure 42 (page 79)
shows the wash logic as a flow diagram. The automatic prism wash
cycle described below consists of three phases: precondition, wash and recovery. The
optional preconditioning function is used to, for example, blow out the condensate before
washing. After the preconditioning there is a one-second pause to avoid having both
precondition and wash relays active at the same time.
The wash cycle is initiated when the wash interval has elapsed. The wash can also be started
by closing an external switch, see
the user interface at the Sensor status display, see
of priority for these wash triggers is as follows:
1. Manual wash
2. Remote wash request
3. Wash interval timer
PRECOND.
WASH
RECOVERY
INTERVAL
Figure 41 Automatic prism wash cycle
Configuring input switches (page
Testing prism wash (page
Chapter 7 – Configuration and calibration
Configuring relays
68), or manually from
51). The order
77