Calibration Check; Testing Prism Wash; Using Indicating Transmitter - Vaisala PR-23 Series User Manual

Process refractometer
Hide thumbs Also See for PR-23 Series:
Table of Contents

Advertisement

6. For a connected sensor, the diagnostic message at start-up is Normal operation or NO
SAMPLE if the process pipe is empty. If another diagnostic message is shown instead,
see
Diagnostic message priorities (page
7. The TEMP value shows the current process temperature.
8. To check the value and the correct setup of the two mA output signals, select
DESCRIPTION in the Main menu and then mA OUTPUTS in the Description menu, see
Viewing system information (page
9. If internal relays or switch inputs are used, you can check their settings through the
Description menu, see
6.1.2 Calibration check
Wait until normal process conditions occur. The concentration reading is precalibrated at
delivery and a copy of the Sensor calibration certificate is inside the Indicating transmitter. If
the diagnostic message is Normal operation but the concentration reading does not agree
with the laboratory results, see
6.1.3 Testing prism wash
1. Check that the steam or water washing parts are connected.
2. In the Main display, press MENU. Then press 3 (to give the command SENSOR STATUS).
In this sensor status display by pressing the soft key WASH. If soft key WASH does not
appear, no internal relay is configured for this purpose.
3. Check the n
reading, for a successful wash it must drop below 1.34 during steam wash
D
and drop to approximately 1.33 during water wash.
CAUTION!
of the refractometer sensor.
More information
‣ Prism wash systems (page 39)
6.2 Using indicating transmitter
The indicating transmitter DTR receives the refractive index value n
temperature from the sensor. Starting from these values, it calculates the concentration of
the process media for display and further transmission. The DTR can also be programmed to
give alarm for high or low concentration. If the refractometer has a prism wash system, the
DTR can control the wash with its built-in timer.
More information
‣ Configuration and calibration (page 59)
97).
55).
Viewing system information (page
Calibrating concentration measurement (page
Before testing prism wash, check that there is liquid in the pipe in front
Chapter 6 – Startup and use
55).
72).
and the process
D
51