IM-EN-PR43IAAX-A Instruction Manual Vaisala K-PATENTS® Process Refractometer PR-43-...-IA/AX/CU...
Page 2
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland P.O. Box 26, FI-00421 Helsinki, Finland +358 9 8949 1 Visit our Internet pages at www.vaisala.com. General safety considerations The process medium may be hot or otherwise hazardous. Use shields and protective clothing adequate for the process medium - do not rely on avoid- ance of contact with the process medium.
Page 3
Note. Notes contain additional information and hints. ® This product manual is delivered to the end user with a Vaisala K-PATENTS product. Information in this manual is subject to change without notice. When the manual is changed, a revised copy is published at http://www.kpatents.com/ Warranty For standard warranty terms and conditions, see www.vaisala.com/warranty.
Page 4
Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
Zone 1: An area in which an explosive gas-air mixture is likely to occur in normal operation. The Vaisala K-PATENTS® Intrinsically Safe Process Refractometer PR-43-...-IA can be in- stalled in hazardous locations in Zone 0 and Zone 1 areas. The PR-43-...-IA refractome-...
Page 8
“n” Note: Servicing of the PR-43-...-IA and PR-43-...-AX/CU refractometers is only allowed for trained service personnel of Vaisala and its representatives. Servicing must be done according to separate instructions defined by Vaisala and must be reported to Vaisala. Document/Revision No.
PR-43-...-IA/AX/CU instruction manual 2.2 PR-43-...-AX/CU refractometers for potentially explosive atmosphere The Vaisala K-PATENTS® refractometer system for potentially explosive atmosphere locations consists of a modified refractometer PR-43 and a connecting cable. The ATEX/IECEx and CSA certified refractometers are identified by the refractometer name- plate, see Figure 2.3 and Figure 2.4 below.
Choose the mounting location of the refractometer and the isolator so that they are protected from sudden impact and friction. If any of the system parts is affected by a sudden impact, power it off immediately and have it checked by trained Vaisala service personnel before it is used again.
Page 14
PR-43-...-IA/AX/CU instruction manual Figure 3.3 Wiring PR-43-...-IA with isolator and PR-8435 cable for Ethernet, mA output and power Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
Page 16
PR-43-...-IA/AX/CU instruction manual Figure 3.5 Wiring PR-43-...-IA with isolator and PR-8432 cable for mA output and power Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
Page 18
PR-43-...-IA/AX/CU instruction manual Figure 3.7 Wiring PR-43-...-IA with isolator, PR-8432 cable for mA and power, and a separate Ethernet cable Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
Page 20
PR-43-...-IA/AX/CU instruction manual Figure 3.9 Refractometer cabling when using Compact user interface CI Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
PR-43-...-IA/AX/CU instruction manual 3.2 Mounting PR-43-...-AX system The refractometer can be installed in a hazardous area up to Zone 2. Use either platform cable PR-8430 or split cable PR-8431 as needed to connect the refractometer e.g. to a DCS. Zone 2 Unclassified area PR-43 sensor Ethernet (M12 connector)
Page 24
PR-43-...-IA/AX/CU instruction manual Figure 3.13 Installing a PR-43-...-AX system with a CI Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
PR-43-...-IA/AX/CU instruction manual Important: Tampering and replacement with other than Vaisala K-PATENTS® origi- nal components is not allowed because this may affect adversely the safe use of the system. Warning! Refractometer connector shall not be connected or disconnected when the circuits are energized.
Page 28
PR-43-...-IA/AX/CU instruction manual Unclassified area Class I Div. 2/Zone 2 Figure 3.16 Cabling in a conduit tube Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
Page 30
PR-43-...-IA/AX/CU instruction manual - La substitution de composants peut rendre ce AVERTISSEMENT! - RISQUE D’EXPLOSION matériel inacceptable pour les emplacements de Classe I, Division 2. Document/Revision No. Effective: October 1, 2019 IM-EN-PR43IAAX...
– EMC requirements – intended for use in industrial locations Notified body Eurofins Expert Services Oy (number 0537) has issued EU-type examination certificate EESF 19 ATEX 056X for this product. Signed for and on behalf of Vaisala Oyj, in Vantaa, on 1 September 2019 ______________________________________ Jari-Pekka Mörsky...
– EMC requirements – intended for use in industrial locations Notified body Eurofins Expert Services Oy (number 0537) has issued EU-type examination certificate EESF 19 ATEX 058X for this product. Signed for and on behalf of Vaisala Oyj, in Vantaa, on 1 September 2019 ______________________________________ Jari-Pekka Mörsky...