Figure 142 Optical Images; Figure 143 Optical Image Detection - Vaisala PR-23 Series User Manual

Process refractometer
Hide thumbs Also See for PR-23 Series:
Table of Contents

Advertisement

The refractive index n
For most solutions the refractive index increases when the concentration increases. At
higher temperatures the refractive index is smaller than at lower temperatures. From this
follows that the optical image changes with the process solution concentration as shown in
the figure below. The color of the solution, gas bubbles or undissolved particles do not
affect the position of the shadow edge (C).
A
Low concentration
Figure 142 Optical images
The position of the shadow edge is measured digitally using a CCD element and is converted
to a refractive index value n
together with the measured process temperature to calculate the concentration.
V
Figure 143 Optical image detection
changes with the process solution concentration and temperature.
D
C
B
by a processor inside the instrument. This value is used
D
a. Optical image
b. CCD element
c. CCD output
A
C
High concentration
Appendix B – Principle of measurement
B
259