Figure 39 Concentration Calibration Layers - Vaisala PR-23 Series User Manual

Process refractometer
Hide thumbs Also See for PR-23 Series:
Table of Contents

Advertisement

1.
2.
3.
FIELD CALIBRATION
4.
5.
6.
Figure 39 Concentration calibration layers
1
The information from the CCD element and the Pt-1000 temperature element. The
position of the shadow edge, see
CCD and scaled 0 ... 100 %.
2
Sensor calibration: The actual refractive index n
process temperature is calculated from the Pt-1000 resistance. The sensor output is n
and temperature TEMP in degrees Celsius. The calibrations of all sensors are identical,
which makes sensors interchangeable. The calibration of each sensor can be verified
using standard refractive index liquids.
3
Chemical curve: The indicating transmitter DTR receives n
concentration value according to chemical curves. The result is a temperature-
compensated calculated concentration value CALC.
CCD
n
D
CHEMICAL CURVE
CALC
DAMPING
CONC
mA OUTPUT
Figure 143 (page
Chapter 7 – Configuration and calibration
Pt-1000
TEMP
259), is described by a number called
is calculated from the CCD value. The
D
and TEMP and calculates the
D
Sensor
Indicating
transmitter
DTR
D
73