Figure 34 Effect of tolerance time on output
7.2.4 QF threshold
The QF threshold setting can be used to prevent the instrument from measuring when the
image quality is below a certain limiting value. When QF value lower than the defined value,
the image status changes to NO OPTICAL IMAGE after the defined tolerance time, see
Tolerance time (page
63).
By default the QF threshold value is -500.
7.2.5 Hold source interactions
There are three reasons why the measurement signal may be in hold. All three result in the
same behavior, but they also interact with each other.
Wash-related hold, see
Hold during wash (page
63), and external hold, see
External hold
(page
63), are connected in parallel. If at least one of them is active, the measurement result
will be in hold.
Tolerance time, see
Tolerance time (page
63), is independent of these two, but the tolerance
time is reset whenever there is another reason for the measurement hold. For example, if the
tolerance time is set to 10 s, and wash hold becomes active after 7 s, the remaining tolerance
time resets to 10 s After the wash is over, there is still 10 s of tolerance time remaining.