Aps/Aps Pro Assy - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-28

APS/APS Pro Assy

NOTE:
APS Power Supply can not be adjusted.
8-2-28-1
Tools
Common pilIips screwdrivers
Nipper
8-2-28-2
Preparations
Shut down the system and switch off the main breaker.
8-2-28-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
2.) Remove the EMI Cover L. Refer to the
3.) Remove Hard Disk assy to unplug the APS/APS Pro connectors on backplane easily.
-
4.) Unscrew 4 screws of the HDD bracket.
5.) Unplug the DC power connector from DVD RW drive.
6.) The APS/APS Pro assy have 5 connectors, 3 connectors are pluged on backplane and 1 connector
is connected to nest cooling fan power and 1 connector is AC power connector.
5
Unplug the HDD power connector on the HDD and the SATA connector from the bottom side
of the backplane. Pull out the connector.
Figure 8-106 Remove HDD
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-15 "Side Left Cover" on page
8-2-26 "EMI Cover L" on page
HDD Connector
SATA Connector
LOGIQ™ P5 S
M
EVICE
ANUAL
8-51.
8-68.
8-71