GE LOGIQ P5 Service Manual page 523

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
10-6-8-1
Data Sheet for ECG Leakage Current
The test passes when all readings measure less than the value shown in the table below. Record all
data on the PM Inspection Certificate.
Table 10-20 Maximum Allowance Limit for ECG Leakage Current
Patient Lead to Ground Leakage Current Test
Patient Lead to Lead Leakage Current Test
Table 10-21 Maximum Allowance Limit for ECG Leakage Current
Patient Lead Isolation Current Test
Table 10-22 Typical Data Sheet for ECG Leakage Current
ECG
Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
5
and
Tester
Tester
Polarity
Ground
Switch
Switch
NORM
CLOSED
REVERSE
CLOSED
NORM
OPEN
REVERSE
OPEN
NORM
CLOSED
REVERSE
CLOSED
NORM
OPEN
REVERSE
OPEN
Section 10-6 - Electrical Safety Tests
AC Power
Source
GROUND OPEN
115V
10uA
220/240V
500uA
AC Power
Source
115V
220/240V
Tester Lead Selector
RL
RA
LOGIQ™ P5 S
EVICE
Maximum
Allowance Limit
GROUND
CLOSED
10uA
10uA
Maximum
Allowance Limit
20uA
5mA
LA
LL
M
ANUAL
C
10-23