Assembly procedure (cont'd)
2.) Assemble the LCD locking mechanism.
a.) Put the stopper pin inside the spring and then put them inside the spring holder. Fix it with
b.) Put lock tite 290 on the M16 tap of the spring holder Assy and screw it.
5
spring pin.
Figure 8-36 Assembling the spring holder Assy
Figure 8-37 Screwing the spring holder Assy
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P5 S
M
EVICE
ANUAL