GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
Table 10-2 Customer Care Schedule
Service at Indicated Time
Clean Probes
Clean Probe Holders
Clean Air Filter
Inspect AC Mains Cable
Inspect Cables and Connectors
Clean Console
Clean Monitor and Touch Panel
Inspect Wheels, Casters, brakes and
Swivel Locks
Check Control Panel Movement
Console Leakage Current Checks
Peripheral Leakage Current Checks
Surface Probe Leakage Current Checks
Endocavity Probe Leakage
Current Checks
Transesphongeal Probe Leakage
Current Checks
Surgical Probe Leakage
Current Checks
Measurement Accuracy Checks
Probe/Phantom Checks
Functional Checks
5
Daily
•*
•
Section 10-3 - Maintenance Task Schedule
Per Facilities
Monthly Quarterly
QA Program
•
•
•
•
•
•
•
LOGIQ™ P5 S
M
EVICE
Notes
* or before each use
more frequently depending on
your environment
Mobile Unit Check Weekly
Mobile Unit Check Daily
Mobile Unit Check Daily
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
•
also after corrective
maintenance
ANUAL