GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
7-5-4-5
APS/APS Pro
These programs are provided for testing the APS board.
Table 7-6
Menu
Full Test
HV voltage
stability
LV voltage
stability
7-5-4-6
FRONT PANEL
These programs are provided for testing the Front panel.
Table 7-7
Menu
FRONT PANEL
Overall
FRONT PANEL
Revision
ROM
RAM
EEPROM
SWITCH
Echoback
NOTE:
For the EchoBack test, it is highly recommended to press a button less than two times per a
second.
NOTE:
When Abort button is pressed to finish EchoBack test, you have to wait until "Completed."
message is displayed to proceed other tests.
7-20
5
FRU Test, APS Diagnostics Menu
Performs all of the APS test listed below
Read HV voltage and display HV values.
Read HV voltage and display LV values.
FRU Test, ACWD Diagnostics Menu
Performs all of the Front Panel tests listed below
Test EEPROM.Send test pattern to write to EEPROM and read.Displays the board
revision of the Front Panel.
Makes a check of internal ROM check sum.
Writes / reads data over all address of RAM and also makes a check of control
program.
Writes / reads data over all address of EEPROM
Test switch in Front Panel.
Rotate or depress any key, Verify that it is displayed on the CRT monitor. To finish the
test, press "Abort" key and wait until complete message appears.
Section 7-5 - Common Diagnostics
Descriptions
Descriptions