Battery On Sysconpm/Cm Assy - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-36

Battery on SYSCONPM/CM Assy

NOTE:
If the battery on SYSCONPM/CM Assy runs out, the system date will be reset. Then can not
exit to maintenance mode.
NOTE:
The Battery is UL critical component. So must use the battery approved. The recommended
one is CR2450 manufactured by TOSHIBA.
8-2-36-1
Tools
Common pilIips screwdrivers
Option strings
8-2-36-2
Preparations
Shut down the system and switch off the main breaker.
8-2-36-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
2.) Remove the EMI Cover R. Refer to the
3.) Unplug the J13 connector. Refer to the Figure 8-123.
4.) Remove the SYSCONPM(SYSCONCM) assy in slot #3 in the nest box.
5
Figure 8-123 Pull out the SYSCONPM(SYSCONCM) Assy
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-16 "Side Right Cover" on page
8-2-27 "EMI Cover R" on page
Unplug the J13 Connector
LOGIQ™ P5 S
M
EVICE
ANUAL
8-52.
8-69.
8-89