Removal procedure (cont'd)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-49
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-9-12
Section 10-5-2-2
Section 10-5-5
Section 10-6-2
Section 10-6-3
Section 10-6-4
Section 10-6-5
Section 10-6-6
Section 10-6-7
Section 10-6-9
8-2-48-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8-110
5
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
ACPC Assy set Function Validation Procedure
Peripheral/Option Checks
Physical Inspection
GEMS Leakage Current Limits
Outlet Test - Wiring Arrangement - USA &
Grounding Continuity
Chassis Leakage Current Test
Isolated Patient Lead (Source) Leakage–Lead
Isolated Patient Lead (Source) Leakage–Lead
Probe Leakage Current Test
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
5394141, Rev 1+, Section 8-2-48. Equipment
Canada
passed all required tests and is ready for use. "