Global Service User Interface (Gsui) - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
5-10-2

Global Service User Interface (GSUI)

5-10-2-1
Internationalization
The user interface provided by the service platform is designed for GE personnel and as such is in
English only. There is no multi-lingual capability built into the Service Interface.
5-10-2-2
Service Login
1.) Touch Utility > Service . It will take about 1p seconds for activations.
2.) Make sure that CAPS is OFF for password entry performed later.
3.) The Service Login window for Service Platform will be shown on the monitor display.
5-10-2-3
Access / Security
The service interface has different access and security user levels. Each user is only granted access to
the tools that are authorized for their use.
NOTE:
A Service Dongle is necessary for use by GE Service when performing proprietary level diagnostics.
OnLine Center access to the scanner requires the password and they must have 'Disruptive'
permission and customer input to run diagnostics.
User Level
Operator
Administrator
External Service
GE Service
NOTE:
For a GE Field Engineer, the password changes at specific intervals. Access with the password is tied
to the service key.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
5
Figure 5-39 Login Screen for Global Service User Interface
Table 5-5
Service Login User Levels
Authorized access to specified diagnostics, error logs and utilities. Same
acquisition diagnostic tests as GE Service.
Knowledge of a service level password.
A physical Service Key (Dongle) required
Section 5-10 - Service Platform
Access Authorization
LOGIQ™ P5 S
M
EVICE
ANUAL
Password
uls
uls
gogems
rotating security
password
5-41