GE LOGIQ P5 Service Manual page 347

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-44-3
Removal procedure (cont'd)
6.) Unscrew the 4 screws on the front of the DVD drive and remove the DVD Assy.
7.) Unscrew 8 screws to remove the bracket from the DVD Assy.
Figure 8-136 Unscrew 8screws and remove the DVD Drive from the bracket
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-45
Service Manual
Section
Section 4-3-1
Section 4-3-2
8-2-44-4
Mounting Procedure
Install the new parts in the reverse order of removal.
5
Figure 8-135 Unscrew 4screws and remove the DVD Assy
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P5 S
EVICE
Debrief Script
"Service Manual, Direction
5394141, Rev 1+, Section 8-2-44. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8-101

Advertisement

Table of Contents
loading

Table of Contents