MAC 1200 V 1.2
Page 33
Information Technologies
227 492 21 RevF
3.4
Keyboard test
By pressing the "2" key the following menu appears:
Key
Test
B=
Bleeper test
E=
End
When a key is pressed this key or its function is displayed. This can either occur by a triple
character display, e.g., "AAA" or as a text, e.g., "CURS_UP". Moreover, pressing the "B" key
tests the bleeper (audible sound). The keys "ALT" and "SHIFT" only in combination with
another key. The "E" key terminates this test and simultaneously undergoes a self-test.
3.5
Motor test
When the "3" key is pressed the following menu appears:
Motor Speed Test
-
Speed selection key
-
Stari/Stop key
Terminate with any key
The speed select key is used to set the required speed and the motor is set into motion with
Start/Stop.
A marking pulse is set once every second. The running speed can then be calculated or its
accuracy checked from the distance between marking pulses. Any other key pressed not
included in this menu leads to the initial display menu.