GE LOGIQ P5 Service Manual page 227

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
7-5-4-3
P3RLY/P2RLY
These programs are provided for testing the P3RLY/P2RLY board.
NOTE:
A Convex Probe is needed for RLY Channel Test.
Table 7-4
Menu
Full Test
Assy Revision
Test
FEBUS Test
SHV Test
OPEN Test
PROBE CODE
7-5-4-4
ACWD
These programs are provided for testing the ACWD board.
Table 7-5
Menu
Full Test
Assy Revision
TEst
USC Interface
Test
FEBUS Test
IQ Data Read
Probe
Connection
Channel Test
5
FRU Test, P3RLY Diagnostics Menu
Performs all of the P3RLY/P2RLY board tests listed below
Test EEPROM.
Send test pattern to write to EEPROM and readDisplays the board revision of the
P3RLY/P2RLY board
Checks if the DSP in SYSCONPM/SYSCONCM can access the P3RLY/P2RLY board
via FE bus.
Send the command to SHV resister in P3RLY/P2RLY board to read SHV status.
Checks whether probe connected or not on P3RLY/P2RLY.
Checks if the 8 bits of the Probe Code in the probe connector is properly connected.
If connected probe is available, display code and Probe name.
FRU Test, ACWD Diagnostics Menu
Performs all of the ACWD board test listed below
Test EEPROM.Send test pattern to write to EEPROM and readDisplays the board
revision of the ACWD board
Checks if the DSP in SYSCONPM/SYSCNCM can access the ACWD boards via USB
bus.
Allows SYSCONPM/SYSCNCM to access FE bus interface in the ACWD in order to
check if FE bus line is normal.
Send the command to perform CW scanning, read IQ data from assigned memory,
analyze the data with FFT algorithm.
Checks whether ACWD probe connected or not.
Generate test pattern, perform 32 channel scanning in ACWD, read the data and
analyze it.
Section 7-5 - Common Diagnostics
Descriptions
Descriptions
LOGIQ™ P5 S
M
EVICE
ANUAL
User
Level
Tools
GE FE
None
Only
All
None
GE FE
None
Only
GE FE
None
Only
GE FE
Probe
Only
User
Level
Tools
All
None
All
None
All
None
7-19