Platform and return to the scan panel.
8-4-4-5
Setting the VCR (if necessary)
1.) Select Utility > System > Peripheral tab.
2.) Under Video Settings, PAL or NTSC.
3.) Under VCR, select Composite or S-Video.
4.) Under VCR, select Panasonic LQ-MD800 or Mitsubishi MD3000, as appropriate.
5.) Click on Save.
6.) Shut down the scanner, then restart the scanner.
7.) Verify that the VCR can be properly operated.
Be sure to connect the converter and cable before booting up the system. Otherwise the LOGIQ™ P5
NOTICE
system does not recognize the peripheral devices