GE LOGIQ P5 Service Manual page 401

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
Platform and return to the scan panel.
8-4-4-5
Setting the VCR (if necessary)
1.) Select Utility > System > Peripheral tab.
2.) Under Video Settings, PAL or NTSC.
3.) Under VCR, select Composite or S-Video.
4.) Under VCR, select Panasonic LQ-MD800 or Mitsubishi MD3000, as appropriate.
5.) Click on Save.
6.) Shut down the scanner, then restart the scanner.
7.) Verify that the VCR can be properly operated.
Be sure to connect the converter and cable before booting up the system. Otherwise the LOGIQ™ P5
NOTICE
system does not recognize the peripheral devices
5
Figure 8-212 Service Platform screen
Figure 8-213 Utility > System>Peripherals
Section 8-4 - Software Installation
LOGIQ™ P5 S
M
EVICE
ANUAL
8-155

Advertisement

Table of Contents
loading

Table of Contents