GE LOGIQ P5 Service Manual page 131

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
4-9-3
Probe Assessment Diagnostic Test Validation Procedure
NOTE:
This diagnostic test does not represent the result as Passed or Failed.
1.) Clean probe surface that you want to assess and connect the probe to the 1st probe slot of the
system.
2.) Turn on the system and check the completion of Echoloader loading.
3.) Do the Service Login.
4.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU Test' => 'Probe' => 'Probe Assesement'.
5.) Execute the diagnostic test.
6.) The test shows bar-graph designating the signal strength of each of the testing probe's elements
including the probe code, name and the number of elements.
5
Figure 4-44 Probe Assessment Diagnostic Test
Section 4-9 - Board Function Checks
LOGIQ™ P5 S
M
EVICE
ANUAL
4-45

Advertisement

Table of Contents
loading

Table of Contents