Mount the Venue 50 to the docking station/cart. Connecting AC/
DC to the docking station/cart, involves preliminary checks of
the power cord, voltage level and compliance with electrical
safety requirements.
1. Ensure that the wall outlet is of appropriate type.
2. Uncoil the power cable, allowing sufficient slack so that the
docking station/cart can be moved slightly.
3. Verify that the power cable is without any visible scratches
or any sign of damage.
4. Verify that the on-site mains voltage is within the limits
indicated on the rating label on the system.
5. Connect the Power Cable to the Power Inlet in the docking
station/cart.
Figure 4-1. Docking Station AC/DC Connection
6. Connect the Power Cable's other end to a hospital grade
mains power outlet with the proper rated voltage, and the
unit is ready for Power ON/Boot Up.
General procedures
4-5