GE Venue 50 Basic Service Manual page 149

Hide thumbs Also See for Venue 50:
Table of Contents

Advertisement

Probe/Connectors Check
NOTE:
Connecting a Probe
Activating the probe
Deactivating the probe
Disconnecting the probe
WARNING
WARNING
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Probes can be connected at any time, whether the unit is ON or
OFF
See 'Connect a probe' on page 3-25 for more information.
The probe activates in the currently-selected operating mode.
The probe's default settings for the mode and selected exam are
used automatically.
When the probe is deactivated, it is automatically placed in
standby mode.
1. Press Freeze.
2. Gently wipe the excess gel from the face of the probe.
(Refer to the Basic User Manual for complete probe
cleaning instructions).
3. Carefully put the probe to the probe holder.
Probes can be disconnected at any time. However, the probe
should not be selected as the active probe.
See 'Disconnect a probe' on page 3-27 for more information.
Take the following precautions with the probe cables: Do
not bend. If you have purchased the cart option, be sure
to keep probe cables free from the wheels.
Be careful not to trip on the probe cables if using the
device without the optional cart.
Functional checks
4-21